APL Technology Inc.

APL ANSWERS

Frequently Asked Questions

Practical answers about barcode inspection, standards, calibration, and testing support.

ANSWERS

Frequently Asked Questions

How do ISO/IEC 15415 and ISO/IEC 15416 differ?

ISO/IEC 15416 grades linear barcodes, while ISO/IEC 15415 applies to two-dimensional codes. The two standards use different grading parameters. DPM marks are evaluated separately under ISO/IEC TR 29158.

Can VeriCube inspect DPM codes?

Yes. REA VeriCube supports ISO/IEC TR 29158 and can inspect laser-marked and dot-peened Data Matrix codes on metal and plastic parts, producing a grade report from 4.0 to 0.

How often should a barcode verifier be calibrated?

As an optical metrology instrument, a barcode verifier is generally recommended for annual calibration according to the manufacturer's interval, with records retained for customer and quality-system audits.

What minimum barcode grade is needed for a supplier audit?

The GS1 General Specifications set 1.5 as the minimum general symbol grade. Medical UDI, automotive, and other sectors may require a higher grade; follow the applicable customer contract or industry standard.

How can I contact APL about a testing solution?

Call +86-10-83682408 and discuss your sample type, symbology, and applicable standard with the technical team. We can provide selection guidance and product information.