Choosing Color Measurement and Whiteness Instruments: Whiteness, Color Difference, and Haze

2026-10-19 · Company and product updates · 澳普乐技术团队

The key point: whiteness, color difference, and haze are three separate properties, answering “how white?”, “how different from the reference?”, and “how clearly can you see through a transparent material?” They use different formulas, standards, and instrument conditions — choosing the wrong equipment often means using a reading for one property to answer a question about another.

This article explains the differences, why whiteness values cannot be compared indiscriminately, and the five conditions to clarify before selecting an instrument.

I. Three properties, three questions

QuestionMetricTypical expression
How white is it?Whiteness (WI, or blue-light brightness R457)A single value, such as 92 or 95
How different is it from the reference?Color difference (ΔE, based on color spaces such as L*a*b*)A tolerance such as ΔE≤1
How clearly can you see through a transparent material?Haze / transmittanceHaze %, transmittance %

These properties are not interchangeable: high whiteness does not mean a small color difference, and an acceptable color difference does not guarantee that whiteness meets customer requirements. Haze is a completely different optical quantity.

II. The key point: whiteness is not simply “one number”

Whiteness can be expressed in many ways, and different industries use different methods. Common measures include blue-light brightness and CIE whiteness, alongside Ganz, Hunter, and Berger whiteness.

1) Blue-light brightness (single-band formula): this measures the short-wavelength diffuse reflectance factor at 457nm, expressed as R457. This method is widely used in paper testing, but different standards and measurement conditions produce different results:

StandardConditions and name
ISO 2470-1Integrating-sphere instrument, indoor daylight conditions (illuminant C), measuring R457 → ISO brightness
ISO 2470-2Integrating-sphere instrument, outdoor daylight conditions (illuminant D65), measuring R457 → D65 brightness
GB/T 7974Follows ISO 2470-2 and adds D65 fluorescence brightness (requires a 420nm UV filter; the difference between R457 under UV-calibrated and UV-excluded conditions is D65 fluorescence brightness)
TAPPI T45245°/0° instrument measuring R457
TAPPI T525Integrating sphere (d/0°) measuring R457
GB 8940.145/0 directional reflectance method, illuminant D65, measuring R457

2) CIE whiteness (two-dimensional): accepted by ISO, ASTM, AATCC, DIN, and others, this is used under illuminants D65 and C in textiles, building materials, nonmetallic minerals, paper, and other industries. The accompanying tint index T indicates a greenish or reddish tint (T>0 greenish, T<0 reddish). Common standards include: ISO 11475 (D65/10°, outdoor light), ISO 11476 (C/2°, indoor light), and ASTM E313.

3) Other measures: Ganz whiteness (see GB/T 17749, GB/T 23774, etc.) and Hunter whiteness (see certain building-material and fiber standards).

III. Standards and measurements: establish the measurement first

MeasurementCommon standardsNotes
Paper blue-light brightness (ISO brightness / D65 brightness)ISO 2470-1 / ISO 2470-2, GB/T 7974Distinguish illuminant C from D65
Paper whiteness (45/0 directional reflectance method)GB 8940.1, TAPPI T452Different instrument geometry
Fiber/pulp brightnessTAPPI T452 / T52545/0 and d/0 configurations
CIE whiteness (paper and board)ISO 11475 (D65/10°), ISO 11476 (C/2°), ASTM E313Used with tint index T
Diffuse reflectance factor (underlying quantity)ISO 2469The underlying measurement for whiteness/brightness calculations
Color calculation and color differenceCIE 15, ASTM E1164, ASTM E308Instrument and calculation procedures
Color of plasticsASTM D6290—
Textile color/color differenceISO 105-J01, AATCC—
Haze and transmittance of transparent filmsASTM D1003, GB/T 2410Requires transmission measurement conditions
Temperature and humidity conditioningISO 187 / TAPPI T402Generally applicable to paper and board

IV. Matching requirements to models

RequirementModelStandards listed on this website
Fluorescent whiteness and brightness of paper/pulpDatacolor Elrepho 1000 fluorescent whiteness testerISO 2469
Benchtop spectrophotometry: whiteness, color difference, and chromaticity (d/8° integrating sphere)Datacolor DC1000 spectrophotometerInstrument: CIE 15, ASTM E1164; software supports ASTM E308, ASTM D6290, ISO 105-J01, AATCC, ASTM D1003 (only supported by DC1000X/1050 transmission versions), ISO 2469, TAPPI
Haze and transmittance of transparent filmsNHM-500 haze meterASTM D1003, GB/T 2410
Full color and appearance testing rangeColor and appearance testing—

V. Five conditions to clarify before selection

  1. Instrument geometry: d/8° integrating sphere or 45°/0°? The same sample gives different values with these geometries, especially for whiteness;
  2. UV handling: for samples containing fluorescent whitening agents (FWA), confirm the UV-calibrated / UV-excluded conditions — this is why D65 fluorescence brightness is calculated by subtracting results under two conditions;
  3. Illuminant and observer: D65/10° or C/2°? Different specifications produce different results;
  4. Aperture and sample quantity: small samples, curved surfaces, and films require appropriate measurement apertures and fixtures;
  5. Sample preparation and conditioning: whether powders are pressed into pellets, the number of paper layers, and whether temperature and humidity conditioning follows ISO 187 / TAPPI T402.

VI. Common mistakes

  1. Comparing whiteness values from different formulas as the same metric (blue-light brightness and CIE whiteness values cannot be compared directly);
  2. Checking whiteness without checking tint index T — at the same whiteness value, greenish and reddish tints differ visually and in subsequent processing;
  3. Ignoring UV conditions — samples with whitening agents give markedly different results with and without UV;
  4. Ignoring geometry differences — 45/0 and d/8 data cannot be mixed;
  5. Ignoring sample preparation, temperature, and humidity — different powder pellet pressures or numbers of paper layers cause results to vary.

VII. Instrument selection and testing support

Send us your sample type + customer-required standard/formula + tolerance, and we will recommend a configuration based on the standards and instrument conditions: