The key point: whiteness, color difference, and haze are three separate properties, answering “how white?”, “how different from the reference?”, and “how clearly can you see through a transparent material?” They use different formulas, standards, and instrument conditions — choosing the wrong equipment often means using a reading for one property to answer a question about another.
This article explains the differences, why whiteness values cannot be compared indiscriminately, and the five conditions to clarify before selecting an instrument.
I. Three properties, three questions
| Question | Metric | Typical expression |
| How white is it? | Whiteness (WI, or blue-light brightness R457) | A single value, such as 92 or 95 |
| How different is it from the reference? | Color difference (ΔE, based on color spaces such as L*a*b*) | A tolerance such as ΔE≤1 |
| How clearly can you see through a transparent material? | Haze / transmittance | Haze %, transmittance % |
These properties are not interchangeable: high whiteness does not mean a small color difference, and an acceptable color difference does not guarantee that whiteness meets customer requirements. Haze is a completely different optical quantity.
II. The key point: whiteness is not simply “one number”
Whiteness can be expressed in many ways, and different industries use different methods. Common measures include blue-light brightness and CIE whiteness, alongside Ganz, Hunter, and Berger whiteness.
1) Blue-light brightness (single-band formula): this measures the short-wavelength diffuse reflectance factor at 457nm, expressed as R457. This method is widely used in paper testing, but different standards and measurement conditions produce different results:
| Standard | Conditions and name |
| ISO 2470-1 | Integrating-sphere instrument, indoor daylight conditions (illuminant C), measuring R457 → ISO brightness |
| ISO 2470-2 | Integrating-sphere instrument, outdoor daylight conditions (illuminant D65), measuring R457 → D65 brightness |
| GB/T 7974 | Follows ISO 2470-2 and adds D65 fluorescence brightness (requires a 420nm UV filter; the difference between R457 under UV-calibrated and UV-excluded conditions is D65 fluorescence brightness) |
| TAPPI T452 | 45°/0° instrument measuring R457 |
| TAPPI T525 | Integrating sphere (d/0°) measuring R457 |
| GB 8940.1 | 45/0 directional reflectance method, illuminant D65, measuring R457 |
2) CIE whiteness (two-dimensional): accepted by ISO, ASTM, AATCC, DIN, and others, this is used under illuminants D65 and C in textiles, building materials, nonmetallic minerals, paper, and other industries. The accompanying tint index T indicates a greenish or reddish tint (T>0 greenish, T<0 reddish). Common standards include: ISO 11475 (D65/10°, outdoor light), ISO 11476 (C/2°, indoor light), and ASTM E313.
3) Other measures: Ganz whiteness (see GB/T 17749, GB/T 23774, etc.) and Hunter whiteness (see certain building-material and fiber standards).
III. Standards and measurements: establish the measurement first
| Measurement | Common standards | Notes |
| Paper blue-light brightness (ISO brightness / D65 brightness) | ISO 2470-1 / ISO 2470-2, GB/T 7974 | Distinguish illuminant C from D65 |
| Paper whiteness (45/0 directional reflectance method) | GB 8940.1, TAPPI T452 | Different instrument geometry |
| Fiber/pulp brightness | TAPPI T452 / T525 | 45/0 and d/0 configurations |
| CIE whiteness (paper and board) | ISO 11475 (D65/10°), ISO 11476 (C/2°), ASTM E313 | Used with tint index T |
| Diffuse reflectance factor (underlying quantity) | ISO 2469 | The underlying measurement for whiteness/brightness calculations |
| Color calculation and color difference | CIE 15, ASTM E1164, ASTM E308 | Instrument and calculation procedures |
| Color of plastics | ASTM D6290 | — |
| Textile color/color difference | ISO 105-J01, AATCC | — |
| Haze and transmittance of transparent films | ASTM D1003, GB/T 2410 | Requires transmission measurement conditions |
| Temperature and humidity conditioning | ISO 187 / TAPPI T402 | Generally applicable to paper and board |
IV. Matching requirements to models
| Requirement | Model | Standards listed on this website |
| Fluorescent whiteness and brightness of paper/pulp | Datacolor Elrepho 1000 fluorescent whiteness tester | ISO 2469 |
| Benchtop spectrophotometry: whiteness, color difference, and chromaticity (d/8° integrating sphere) | Datacolor DC1000 spectrophotometer | Instrument: CIE 15, ASTM E1164; software supports ASTM E308, ASTM D6290, ISO 105-J01, AATCC, ASTM D1003 (only supported by DC1000X/1050 transmission versions), ISO 2469, TAPPI |
| Haze and transmittance of transparent films | NHM-500 haze meter | ASTM D1003, GB/T 2410 |
| Full color and appearance testing range | Color and appearance testing | — |
V. Five conditions to clarify before selection
- Instrument geometry: d/8° integrating sphere or 45°/0°? The same sample gives different values with these geometries, especially for whiteness;
- UV handling: for samples containing fluorescent whitening agents (FWA), confirm the UV-calibrated / UV-excluded conditions — this is why D65 fluorescence brightness is calculated by subtracting results under two conditions;
- Illuminant and observer: D65/10° or C/2°? Different specifications produce different results;
- Aperture and sample quantity: small samples, curved surfaces, and films require appropriate measurement apertures and fixtures;
- Sample preparation and conditioning: whether powders are pressed into pellets, the number of paper layers, and whether temperature and humidity conditioning follows ISO 187 / TAPPI T402.
VI. Common mistakes
- Comparing whiteness values from different formulas as the same metric (blue-light brightness and CIE whiteness values cannot be compared directly);
- Checking whiteness without checking tint index T — at the same whiteness value, greenish and reddish tints differ visually and in subsequent processing;
- Ignoring UV conditions — samples with whitening agents give markedly different results with and without UV;
- Ignoring geometry differences — 45/0 and d/8 data cannot be mixed;
- Ignoring sample preparation, temperature, and humidity — different powder pellet pressures or numbers of paper layers cause results to vary.
VII. Instrument selection and testing support
Send us your sample type + customer-required standard/formula + tolerance, and we will recommend a configuration based on the standards and instrument conditions: