White paper—July 2025
REA PRINT | APPLY | VERIFY
The science of Unique Device Identification (UDI) Marking: practical applications and insights
Authors Wilfried Weigelt, Head of Department - Barcode Verifier, REA Tobias Thiel, Business Development Manager Pharmaceutical and Medical Care, REA Nicole Kaller, Application Development Leader EMEA, Tyvek® Healthcare Packaging, DuPont
Executive summary
The Unique Device Identification (UDI) system enables clear identification of specific medical devices available on the market and enhances their traceability. UDI information must be presented in both human‑readable and machine‑readable formats, with print quality adhering to the relevant application specifications and ISO standards.
Manufacturers have access to various marking and printing systems and must choose the most suitable technology for their specific applications. Additionally, a quality assurance process must be established to ensure the accurate readability of codes, meeting legal requirements as well as application specification demands referenced (indirectly) by law.
This paper provides background information on the UDI system and discusses several printing and marking technologies used for variable data, as well as the quality assurance (grading) processes for printed codes. It also covers related issuing entities such as GS1, and standards such as ISO/IEC 15415 applicable for 2D matrix codes. This ISO standard has been revised in 2024, which introduces significant changes, including the implementation of decimal grading instead of the integer grading used in the previous release. This change is expected to enhance the reliability of grading results.
A printing and grading study was conducted using DuPont™ Tyvek® which demonstrated that good print quality could be achieved across all three Tyvek® grades—1073B, 1059B and 2FS—utilizing all REA inkjet technologies (water‑based, solvent‑based and UV‑based). The grading process successfully exceeded quality requirements with the use of REA VeriCube (offline grading) and VeriMax (inline grading). Grading according to ISO/IEC 15415:2024 provided more accurate and better results than according to ISO/IEC 15415:2011. The study also highlighted that barcode or print quality is influenced by factors such as ink selection, print settings (including resolution), pixel reduction and barcode size.
For more detailed information geared towards printing and grading professionals, the annexes of this document provide valuable insights into printing technologies, the revision of ISO/IEC 15415, differences between scanners and code measuring instruments (verifiers), as well as the benefits of inline quality control.
Unique Device Identification (UDI) ‑ Introduction
Healthcare personnel often encounter challenging situations when managing medical devices, such as: “Is this the correct device for this patient?” “How old are our reusable devices?” “Is the traceability and usage information for the medical device accessible to users as required by law?” “Where should I file the device information?” ...and many others.
Unique Device Identification (UDI) offers a solution to these challenges. The UDI is a distinct numeric or alphanumeric code associated with a medical device, enabling clear identification of specific medical devices available on the market and facilitating their traceability. UDI is designed to provide critical information about the origin and characteristics of a medical device throughout its life cycle.
The fundamental requirements and guidelines for UDI were established by the International Medical Device Regulators Forum (IMDRF [3]), aiming to harmonize UDI requirements on a global scale.
The UDI data for each medical device must be uploaded to region‑specific databases, such as the Global Unique Device Identification Database (GUDID [1]) in the United States, the Eudamed [2] database in Europe, and the Unique Device Identification Database (UDID [27]) in China.
While the implementation of UDI entails additional efforts for manufacturers, it also presents opportunities to restructure and enhance processes and efficiency in medical device manufacturing and logistics.
Europe/EU: According to the Delegated Act 2017/745 [4], medical devices sold in the EU should be uniquely identifiable by the end of December 2028. The deadlines as per the Medical Devices Regulation (MDR) depend on whether medical devices are classified as high, medium or low risk. High‑risk devices (e.g. artificial joints or heart valves) had to be UDI‑labeled by May 2021, in vitro diagnostic devices by May 2022. For the other classes, the deadlines have been extended. Depending on the type of device, medium‑risk (e.g., anesthesia or hearing aids) and low‑risk (e.g., walkers and bandages) medical devices must be UDI‑labeled at latest by 2025, by the end of December 2027 or by the end of December 2028, respectively.
United States: UDI labeling requirements are defined in the Regulation 21 CFR Part 801. The final rule was published by the U.S. Food and Drug Administration (FDA) in 2013 (58786 Federal Register/Vol. 78, No. 185/Tuesday, September 24, 2013/Rules and Regulations [26]. The enforcement of UDI requirements for all medical devices has been in effect since September 2023.
China: In 2022, the Chinese National Medical Products Administration (NMPA) released regulations for UDI NMPA [27]. These regulations encompass the assignment of UDI, the coding that carries the UDI and the UDID. The principles concerning data carriers, the application of international standards and the selection of issuing agencies closely resemble those established in the United States and the European Union. The implementation phase for UDI is anticipated to be completed for all device classes by 2026.
What is UDI?
UDI = DI+PI (Unique Device Identification = Device Identifier + Production Identifier)
Figure 1 (01)04270001108715(11)201224(17)260731 (10)REA‑Lot (21)REA‑Serial(01)04270001108715(11)201224(17)260731(10)REA‑Lot(21)REA‑Serial
UDI‑DI — Device Identifier (static data) ‑ Serves to identify product, manufacturer and risk class ‑ Contains the GTIN (Global Trade Item Number) (GS1), the UPN (Universal Product Number) (HIBCC) or the ISBT 128‑PPIC (Processor Product Identification Code) (ICCBBA), depending on the issuing agency
UDI‑PI — Production Identifier (variable data) ‑ Batch or lot numbers, serial numbers, expiry and / or manufacturing date ‑ Assigned by the manufacturer
A UDI begins with the Device Identifier (DI), which is a globally unique item number created following appropriate ISO standards [5]. Four recognized issuing entities— GS1[7], HIBCC [8], ICCBBA [9] and IFA [10] —establish the procedures to ensure the global uniqueness of the DI [11].
The second component is the Production Identifier (PI) [12] which includes variable data associated with the medical device, such as the production or expiration date, lot or batch number and serial number.
In Europe, UDI information for medical devices must be uploaded to the Eudamed database (European Databank on Medical Devices). Each manufacturer is responsible for registering their medical devices with their respective UDI, while importers, distributors, pharmacies, clinics and practices contribute to and complete the data. The purpose of the Eudamed database is to document the characteristics of each product. The U.S. equivalent of Eudamed is the Global Unique Device Identification Database (GUDID), while in China, it is referred to as the Unique Device Identification Database (UDID).
A typical UDI solution employs machine‑readable codes, such as 1D linear barcodes or 2D matrix codes like QR Codes or Data Matrix codes. These codes are traditionally utilized for automatic identification and for streamlining logistics and production processes.
The primary objective of UDI is to enhance patient safety through improved product traceability. Additionally, compliance with UDI regulatory requirements is expected to boost efficiency throughout the supply chain, from manufacturing to dispensing to the patient.
UDI Code Marking – 1D and 2D Barcodes
To ensure accurate and rapid identification of each medical device, a machine‑readable code—either a 1D linear barcode or a 2D matrix code—is applied to the product. This code contains the UDI data structured in accordance with relevant ISO standards [13]. Practical implementation details are defined by application specifications from issuing agencies such as GS1 or HIBCC, which reference ISO standards.
While traditional 1D linear barcodes are still used, they are often less suitable due to their higher space requirements and lower error tolerance compared to 2D matrix codes. Consequently, 2D matrix codes have become the preferred option.
The machine‑readable code should be accompanied by human‑readable text, provided there is sufficient space on the product for both formats. UDI information is typically printed using on‑demand marking processes due to the variability of the data, and several marking process options are available for manufacturers to choose from.
Printing and Marking — Choosing the right printing technology
The wide range of materials used in medical devices and packaging presents a challenge for effective labeling. To guarantee that the Unique Device Identification (UDI) is both dependable and consistently legible, it is crucial to select the most suitable printing or marking technology.
Thermal transfer (TT) printing is a process in which the heating element of a print head transfers ink from a ribbon directly onto a substrate, such as a label. Depending on the specific combination of substrate and ribbon chosen, this method can produce high‑quality and durable printed images. When printing on Tyvek®, it is crucial to select the appropriate ribbon to ensure optimal results. REA LABEL provides a variety of printers and application systems tailored for this printing technology.
Inkjet printing can be seamlessly incorporated into the production line for packaging serialization. In this method, ink droplets are directed onto the packaging substrate without the print heads making contact with the material, which distinguishes it from processes like thermal transfer printing.
‑ Thermal inkjet (TIJ) printing operates by utilizing a heater in the nozzle to generate a vapor bubble that ejects the ink droplet, leading to the term “thermal inkjet” (TIJ). This technique is particularly well‑suited for DuPont™ Tyvek®, a commonly used sterile packaging material in healthcare applications. Composed of polyolefin filaments, Tyvek® is gas‑permeable, allowing for the sterilization of packaged medical devices, and boasts high tear and puncture resistance. Given the characteristics of the material, it is essential to choose inks that dry quickly and adhere effectively. The REA JET HR system, which employs TIJ (HP) cartridges, offers a broad range of ink types that ensure excellent print quality on Tyvek®.
‑ Piezo inkjet (PI) printing employs a piezoelectric crystal in the nozzle, which rapidly changes shape in response to electrical voltage, allowing it to eject ink droplets onto the printing surface. Industrial printing systems, such as the REA JET UP, offer a substantial print height of approximately 100 mm and deliver a resolution comparable to that of thermal inkjet (TIJ) printers. However, because a lower print height is typically needed for printing UDI on healthcare packaging, piezo inkjet printers are less frequently utilized in this specific context.
‑ Continuous Inkjet (CIJ) systems operate as a single nozzle system, producing a continuous stream of ink droplets. The placement of each droplet is controlled by applying an electrical charge to it and using an electric field to direct its trajectory. While this method allows for quick drying of the ink, the resulting print consists of individual dots that exhibit low resolution, making them not easily discernible to the naked eye. This presents a trade‑off: Although the ink dries quickly, the lower print resolution and accuracy are notable drawbacks. CIJ printers can be utilized on Tyvek®, but REA does not recommend this technology for printing UDI codes on it due to the inadequate resolution and a significant reduction in speed when handling more complex print layouts.
For additional details about various printing technologies, please refer to ANNEX A.
Figure 2 ‑ UDI marking of DuPont™ Tyvek® medical device packaging for patient safety with the REA JET HR TIJ system.
Marking quality of the UDI Code – Barcode grading
Don’t Forget to Check Print Quality High‑quality and legible UDI information is essential for the rapid and accurate identification of defective, recalled or expired medical devices. It is vital for manufacturers to ensure that the marking is correct and free of errors, that the content is precise, and that the quality meets international standards. To assess code quality, specific optical measuring instruments known as verifiers should be used as simple scanners do not have the capability to accurately measure and exhibit significant technological and performance variations (including those found in smartphones, industrial handheld scanners and fixed inline scanners).
In Europe, there are four recognized issuing entities or application specifications for UDI: GS1, HIBCC, IFA and ICCBBA. Among these, GS1 is the most widely recognized, operating in a total of 145 countries, including various nations in Europe and the United States. This broad recognition highlights the importance and global applicability of GS1 standards in the healthcare sector.
All listed entities refer to ISO/IEC 16022 [17] (Automatic identification and data capture techniques — Data matrix bar code symbology specification), ISO/IEC 29158 (Automatic identification and data capture techniques. Direct Part Mark (DPM) Quality Guideline), ISO/IEC 15415 (Automatic identification and data capture techniques. Bar code symbol print quality test specification. Two‑dimensional symbols) and other relevant standards. These standards provide the most appropriate guidelines for evaluating the print quality of UDI codes, specifically in terms of barcode grading.
Figure 3 ‑ DPM verification with the REA VeriCube verifies UDI marking for information and traceability of the medical device throughout its life. Figure 4 ‑ Barcode grading with the REA VeriMax offline verifier.
表格
| Parameter | Meaning |
| Symbol contrast (for DPM: cell contrast) | Difference in reflectance between the dark and light elements of a barcode |
| Modulation and reflectance margin (for DPM: cell modulation) | Refers to the variation in the reflectance of the dark and light bars of the barcode and the uniformity of reflectance of light and dark module areas of a 2D matrix code |
| Fixed pattern damage (FPD) | Inspection of areas in a 2D matrix code that do not contain data, but are used for position recognition, grid reconstruction, etc. |
| Decode | Reference Decode algorithm as defined by the ISO symbology standard, is applied and checked if it can decode successfully. Invalid use of codewords is checked in addition. Further predictable and consistent errors in barcode patterns or elements that result in parts of the barcode not aligning with the standard specifications (Decode definition differs for barcodes and 2D matrix codes). |
| Unused error correction (UEC) | Reports if destroyed code parts were detected and corrected |
| Axial non‑uniformity (AN) | Shows whether a code (and single matrix cells) are stretched or squeezed |
| Grid non‑uniformity (GN) | Shows whether matrix cells are at the correct position within the barcode |
| Module size | Verifies the average matrix cell width (calculated from each single matrix cell). The acceptable range depends on the requirement of an application specification* |
| Print growth | Measures whether matrix cells are printed too large or too narrow |
| Minimum reflectance (for DPM) | Amount of light that is reflected from the light‑colored areas of the barcode when illuminated |
*An application specification defines the Module size as X dimension which is the perfect size of any single matrix cell.
The quality grades are expressed using a numeric scale that ranges from 4.0 (indicating the best quality) to 0 (indicating failure), with increments of 0.1. Each of the parameters mentioned above is evaluated according to this scale.
Figure 6
表格
| New Quality grades (steps of 1/10) | Old integer grades | Historical ANSI‑Grades | Meaning |
| 3.5 to 4.0 | 4 | A | Very good |
| 2.5 to < 3.5 | 3 | B | Good |
| 1.5 to < 2.5 | 2 | C | Satisfactory |
| 0.5 to < 1.5 | 1 | D | Adequate |
| < 0.5 | 0 | E | Failed |
If GS1 has been selected as the issuing entity, the quality requirements outlined in paragraph 5.12.3.x of the GS1 General Specification standard must be followed. GS1 mandates a minimum quality grade of 1.5 for barcodes associated with healthcare products. It is to be noted that the GS1 General Specifications are updated annually (currently at Release 25.0, January 2025).
If a different issuing entity is selected, the specifications set forth by that entity must be adhered to.
Specialized code measurement systems, such as REA Verifier [18] instruments, assist manufacturers in the metrological and qualitative evaluation of their UDI markings. These systems can be stationary, mobile or integrated into production for both offline and inline barcode grading. The quality criteria established by the selected issuing entity are based on legal requirements and must be met. Quality evaluation systems incorporate checks to confirm compliance with these criteria.
For manufacturers of medical devices, it is essential to take proactive steps in the UDI domain and collaborate with an experienced partner. This partnership should encompass all aspects from code development to appropriate labeling methods and code verification solutions.
Revision of ISO/IEC 15415
A revised version of the quality test standard for 2D codes has been published in December 2024. This revision introduces several significant changes aimed at producing more consistent quality measurements, with finer scaling and better reproducibility of results. The following summarizes some of the changes: ‑ Continuous grading, also known as decimal grading, has been introduced. Numeric grading is provided in 41 steps compared to only 5 steps before. The quality grades are expressed on a descending scale from 4.0 to 0 in steps of 0.1. ‑ Improved threshold determination between dark and light. ‑ Grading of too fat or narrow matrix cells (print growth/loss). ‑ Removal of inspection of reflectance outside of the code. ‑ Adoption of the filter to blur the image (synthetic aperture) to current scanner technology. ‑ Combining modulation and reflectance margin into a single parameter: Modulation.
More details can be found in ANNEX C.
Printing & Bar Code Quality Evaluation Study
Study concept The purpose of this study was to confirm the print quality of variable data on DuPont™ Tyvek® Healthcare grades printed using inkjet printing technology from REA. Variable data was printed in the form of a GS1 DataMatrix code. Inkjet technology with water‑, solvent‑ and UV‑curing‑based ink was selected. Offline and inline grading equipment were used to evaluate the quality of the matrix codes.
The study was divided into two parts:
- Pre‑screening trials: These trials aimed to define the most optimal matrix code module sizes, inks and print resolution (with or without pixel reduction) to achieve good print quality. An offline grading system was used.
- Main study: This part applied the learnings and settings from the pre‑screening trials using an inline grading system.
Part 1: Pre‑screening trials The objective of the pre‑screening trials was to find the optimal print resolution and Data Matrix module size for printing on DuPont™ Tyvek® 1073B, 1059B and 2FS™. The minimum requirement was to achieve a grade of at least 1.5, according to ISO/IEC 15415, using the offline grading equipment REA VeriCube.
GS1 DataMatrix: ‑ 3 different module sizes:0.25 mm, 0.45 mm and 0.50 mm
Module size refers to the size of a single matrix cell* *If measured the average of all single matrix cell sizes (x and y direction) is the reported module size
Print resolution: ‑ 300×300 dpi, with and without print optimization (pixel reduction) ‑ 300×260 dpi, with and without print optimization (pixel reduction)
Ink types: ‑ 3 water‑based, 2 solvent‑based and 1 UV curable ink
Figure 8 ‑ Identical codes with different module sizes. Figure 9 ‑ Print layout
Results, findings and conclusions ‑ Module size 0.25 mm: Not all codes could be correctly read or verified with the VeriCube according to ISO/IEC 15415, resulting in a grading of 0. ‑ Module size 0.45 mm: Grades of 1.5/C could be achieved with all print resolutions. In individual cases, grades of 2/C to 3/B were obtained. The best result was achieved with a 300×300 dpi resolution while applying print optimization and pixel reduction. ‑ Module size 0.50 mm: Grades of 1.5/C could be achieved with all print resolutions. In most cases, grades of 3/B to 4/A were obtained with a resolution of 300×300 dpi while applying print optimization and pixel reduction.
Recommendations for print Due to the substrate structure, which is not entirely flat, it is more challenging to achieve the minimum grade of 1.5/C using small module sizes. It is recommended to print the 2D code with a resolution of 300×300 dpi, applying print optimization and pixel reduction. The recommended data matrix module size is 0.50 mm. All different ink types provided good results.
It is to be noted that a good print quality forms the basis for high process reliability.
Part 2: Main study ‑ Inline printing & grading trials During the main study, the optimized settings and 2D Data Matrix module size of 0.50 mm, as defined in the pre‑screening trials, were used for printing on DuPont™ Tyvek® 1073B, 1059B, and 2FS™. The minimum requirement was again to achieve a grade of 1.5/C, according to ISO/IEC 15415. This time, the inline grading equipment REA VeriMax was used at a line speed of approximately 12 m/min.
GS1 DataMatrix: ‑ One module size:0.50 mm Print resolution: ‑ 300×300 dpi, with and without print optimization (pixel reduction) Ink types: ‑ 3 water‑based, 2 solvent‑based and 1 UV curable inks
Figure 10 ‑ Print layout Figure 11 ‑ Variable data print layout proof for the printing study. Figure 12 ‑ REA Inkjet variable data and barcode printing on Tyvek®. Figure 13 ‑ REA VeriMax inline barcode grading of barcodes printed on Tyvek®.
Results, findings and conclusions Grading results ranging from 2.0/C to 4.0/A were achieved on all three DuPont™ Tyvek® products: 1073B, 1059B and 2FS™. All codes met the acceptance criteria, with all results well above the minimum quality grade of 1.5/C. All types of inks used provided good results on all Tyvek® grades.
Figure 14 ‑ Barcode grading results summary (per ISO/IEC 15415:2011 and ISO/IEC 15415:2024)
表格
| Ink type | Ink name | Tyvek® grade | Grade (ISO/IEC 15415:2011) | Grade (ISO/IEC 15415:2024) | Grade >1.5 |
| Water‑based | Black Eagle CS | 1073B | 3 | 3.4 | Pass |
| Water‑based | Black Eagle CS | 1059B | 3 | 3.6 | Pass |
| Water‑based | Black Eagle CS | 2FS | 4 | 4.0 | Pass |
| Water‑based | Black Eagle HD | 1073B | 3 | 3.5 | Pass |
| Water‑based | Black Eagle HD | 1059B | 3 | 3.5 | Pass |
| Water‑based | Black Eagle HD | 2FS | 3 | 4.0 | Pass |
| Water‑based | TWP 627 | 1073B | 2 | 3.3 | Pass |
| Water‑based | TWP 627 | 1059B | 2 | 3.6 | Pass |
| Water‑based | TWP 627 | 2FS | 3 | 4.0 | Pass |
| Alcohol‑based | Multiple black V2 | 1073B | 3 | 3.7 | Pass |
| Alcohol‑based | Multiple black V2 | 1059B | 3 | 3.5 | Pass |
| Alcohol‑based | Multiple black V2 | 2FS | 4 | 4.0 | Pass |
| Alcohol‑based | Multiple black DC V3 | 1073B | 3 | 3.6 | Pass |
| UV curable | UV curable black | 1073B | 3 | 4.0 | Pass |
| UV curable | UV curable black | 1059B | 2 | 3.1 | Pass |
Grade mapping:3.5‑4.0/A;2.5‑<3.5/B;1.5‑<2.5/C;0.5‑<1.5/D;<0.5/F Not measurable
Conclusions of the study High‑quality printing could be achieved on all three DuPont™ Tyvek® grades 1073B, 1059B, and 2FS™ using REA inkjet technologies ‑ water, solvent and UV‑based. A module size of 0.50 mm and a resolution of 300×300 dpi while applying pixel reduction provided the best results.
Barcode grading has been successfully performed, exceeding quality requirements, with the REA VeriCube (offline grading) and VeriMax (inline grading). Barcode or print quality is influenced by ink selection, print settings such as resolution, pixel reduction and barcode (module) size.
Higher resolution results in more ink on the printed material, leading to higher contrast and longer drying times. However, the final print quality also depends on the material’s ink absorption. Excessive ink can cause undried ink or ink bleeding, affecting print quality and grading results, and increasing the risk of ink penetration through the material. Therefore, it’s crucial to adjust the ink amount and resolution for each material used.
Grading results ranging from 3.0/B to 4.0/A were achieved on all tested DuPont™ Tyvek® products and with different types of inks. All codes met the acceptance criteria, with all results well above the minimum quality grade of 1.5/C. Grading according to ISO/IEC 15415:2024 provided more accurate and better results than according to ISO/IEC 15415:2011.
For more information on optimizing your variable printing and grading process, please contact us.
About us
REA As a full‑range supplier, REA Elektronik is ready to support the industry in UDI implementations tasks with its REA JET, REA LABEL and REA VERIFIER product lines. Industry 4.0‑capable systems that can be seamlessly integrated into existing production lines are indispensable for individualized product marking. At the same time, modern labeling processes are required to prepare manufacturers for optimized internal and external workflows.
DuPont DuPont develops solutions that protect and shelter; healthcare, aerospace, industrial and transportation industries. DuPont solves challenges with high‑performance fibers, aramid papers, non‑woven structures, water‑purification technologies and advanced materials. Tyvek® healthcarepackaging.tyvek.com 1 800 44 TYVEK
© 2025 REA Elektronik GmbH
声明段落(版权免责) Nothing contained herein shall be construed as a representation that any recommendations, use or resale of the product or process described herein is permitted and complies with the rules or regulations of any countries, regions, localities, etc., or does not infringe upon patents or other intellectual property rights of third parties. The information provided herein is based on data REA Elektronik GmbH believes to be reliable. Accordingly, REA makes no warranty for its use. If translated, original English version controls.
DuPont trademarks: Tyvek®, Kevlar®, Nomex®, etc are trademarks of affiliates of DuPont de Nemours, Inc.
ANNEX A On‑demand printing technologies for variable information and data
Healthcare packaging is typically partially preprinted with fixed data, such as logos and usage information, while variable data, like UDI information, is printed on demand. Technologies well‑suited for on‑demand printing on DuPont™ Tyvek® include inkjet and thermal transfer. Laser marking, such as laser ablation as used on carton boxes, is critical to avoid because the laser can melt the filaments of Tyvek®.
Applicable inkjet technologies for printing on Tyvek® are: ‑ Thermal inkjet (TIJ bubble Jet) (drop on‑demand) ‑ Piezo inkjet (drop on‑demand) ‑ Continuous inkjet (CIJ) (continuous ink drop stream)
Thermal transfer (TT) Thermal transfer uses a ribbon between the print head and the substrate for transferring the ink to the substrate. This method can print self‑adhesive labels or mark parts directly. Print speeds: 50‑600 mm/s; resolutions: 200 dpi ~ 600 dpi. For Tyvek®, wax‑resin ribbons are needed to fill the fibrous surface.
Thermal inkjet (TIJ) Most popular for UDI on Tyvek®, using HP‑style disposable print cartridges. Speed: 50‑150 m/min for complex Data‑Matrix. Resolution up to 600 dpi. Print‑head‑to‑substrate gap ~1 mm. Solvent‑modified inks are preferred for hydrophobic Tyvek. Multiple print‑heads can be arranged for larger print height.
Piezo Inkjet Permanent piezo print‑head, separate ink supply. REA JET UP achieves ~360 dpi, large print height (~100 mm). Less common for Tyvek healthcare UDI.
Continuous Inkjet (CIJ) High‑speed, large working distance. But low resolution (~50‑100 dpi), round ink‑dots not ideal for sharp matrix‑cell squares. Not recommended for UDI on Tyvek by REA.
ANNEX B Barcode scanning vs. measurements (grading)
Core point: A handheld scanner can “read” a bad code, but cannot perform standardized quality grading. Verifiers (per ISO 15426‑2) use reference‑decode, fixed lighting/aperture, controlled conditions for objective grading. Scanners are optimized for decode‑success, not metrological measurement.
Key distinctions: ‑ Scanner: pass/fail decode; variable lighting/angle; inconsistent results. ‑ Verifier: standardized illumination, aperture, dark background; calibrated hardware; outputs numerical grade for each parameter. ‑ Even codes that “scan fine” may fail in fixed‑mount high‑speed production; verifier catches latent defects early.
Mandatory standards for 2D‑verifier hardware: ISO/IEC 15426‑2. Test‑methods: ISO/IEC 15415 (printed symbols), ISO/IEC 29158 (DPM direct‑part marking). Verifiers must be periodically calibrated with reference test‑cards.
ANNEX C Revision of ISO/IEC 15415 (2024‑12 release)
Major updates versus 2011 edition:
- Decimal grading scale: 0.0‑4.0 in 0.1 increments (total 41 grades), replacing coarse integer grades 0‑4.
- Threshold calculation uses image‑histogram statistics (1% brightest / 1% darkest pixels), rejects single‑pixel extreme reflectance outliers.
- Print‑growth / print‑shrinkage becomes graded parameter: deviation of measured module size vs nominal X‑dimension. <10% error = Grade4; >60% = fail; Grade1.5 allows up to 37% dimensional error.
- Removed the asterisk annotation for out‑of‑symbol background reflectance.
- Synthetic‑aperture filter rules updated; supports aperture sized proportionally to measured module dimension.
- “Reflectance margin” merged into the renamed single parameter Modulation.
ANNEX D Inline Quality Control
Inline verification inspects codes while parts move along production. Simple fixed‑mount scanners only check “code present / content match”. True inline grading (REA VeriMax) must satisfy full ISO‑verifier hardware requirements. Motion blur limits maximum line speed. Inline‑grading is also called pseudo‑measurement in some contexts.
Literature references
[1] US‑FDA GUDID [2] EU Eudamed [3] IMDRF [4] EU MDR Delegated Act 2017/745 [5] ISO/IEC 15459‑3 [6] ISO/IEC 15459‑2 [7] GS1 Healthcare UDI [8] HIBCC [9] ICCBBA / ISBT‑128 [10] IFA‑FFM [11] EU‑Commission UDI guidance [12] medical‑device‑regulation.eu UDI‑PI [13‑27] ISO‑standards, NMPA‑UDID database, REA‑Verifier product documentation, Federal Register FDA UDI final rule.